TA-25NL and TA-45NL are part of a family of calibration baths designed for large calibration volume. This Calibrators family is ideal to calibrate sensors that require deep insertion. They also have HART® Communication Protocol, data logger mode, Ethernet and USB interface.
| Parameter | TA-25NL | TA-45NL |
|---|---|---|
| Operating Range | -25 °C to 140 °C (1) | -45 °C to 140 °C (1) |
| Power Supply | 115 Vac or 230 Vac 50/60 Hz, according to order code. | |
| Well Diameter / Depth | Ø 35 mm x 160 mm | |
| Resolution | 0.01 °C (0.01 °F) | |
| Display Accuracy | ± 0.10 °C | |
| Axial Uniformity (40 mm) |
± 0.06 °C (Dry Block) ± 0.025 °C (Stirred Liquid) |
± 0.08 °C (Dry Block) ± 0.025 °C (Stirred Liquid) |
| Radial Uniformity |
± 0.02 °C (Dry Block) ± 0.02 °C (Stirred Liquid) |
± 0.025 °C (Dry Block) ± 0.02 °C (Stirred Liquid) |
| Stability (after 10 min) | ± 0.02 °C (0.04 °F) | |
| Power Consumption | 300 W (rated) | 450 W (rated) |
| Heating Rate | 25 minutes (25 °C to 140 °C) |
20 minutes (25 °C to 140 °C) |
| Cooling Rate | 20 minutes (25 °C to -25 °C) |
40 minutes (25 °C to -45 °C) |
| Dimensions (H,W,D) | 315 x 200 x 305 mm | |
| Weight | 12 kg | |
| Warranty | 1 year, except for Peltier effect elements | |